Maintenance and calibration service of automatic pipettes mono channel
Pipettes are instruments whose normal use causes inaccuracies over time, therefore it is necessary to make adjustments and calibrations periodically to ensure accuracy and repeatability in each measurement in order to obtain reliable results. We perform cleaning and calibration of all brands and models of automatic pipettes.
Methodology
1. The condition of the pipette is evaluated.
2. All corroded or residual parts are dismantled and cleaned.
3. All parts are greased and assembled.
4. It is calibrated on analytical balance.
Implementing
This service is available to all centers, institutes and laboratories in the region.
Frequently asked questions
How often is maintenance necessary?
This type of volumetric material requires maintenance every 3 months to ensure its accuracy and repeatability.
What does maintenance include?
Depending on the condition of the pipette, we perform cleaning, greasing, adjustment and calibration.
What happens if the pipette is not maintained?
Pipettes are instruments whose normal use causes inaccuracies over time, therefore it is necessary to make adjustments and calibrations periodically to ensure accuracy and repeatability in each measurement in order to obtain reliable results. Without proper maintenance, pipettes can lead to volume measurement errors and even lead to contamination in tests.
Technical Responsible
Francisco Badalini Nanni
2236363151